1 kg
10 °
0 .. 100°C
ambient
1 µ°
ECGp5050/Al/HL/NUM/RT
bearing based goniometer for Φ-positioning
The ECGp5050 provides high-resolution Φ-positioning over a 10° range. Combined with the ECGt5050, the ECGp5050 forms the lower positioner of a powerful Euler Goniometer with a common center of rotation. It can be combined with a xy-stack of ECSx5050 to add two more degrees of freedom to the motion of the sample.
The positioner comes with an integrated optoelectronic encoder (/NUM),with high dynamic force for higher loads (/HL) and is designed for ambient conditions (/RT).
Our aluminum models combine a powerful piezoelectric motor with a lightweight body for easy integration in setups in atmospheric conditions.
Specifications
For a detailed definition of used terms and descriptions please visit our Nanopositioners Glossary
Size and Dimensions | |
---|---|
footprint; height | 50 mm x 50 mm; 17 mm |
distance center of rotation to bottom | 94 mm |
maximum installation space | 50 mm x 69.8 mm; 19.5 mm |
weight (aluminium version) | 137 g |
height | 17 mm |
Materials | |
positioner body | aluminum |
actuator | PZT ceramics |
connecting wires | copper, jacket: RT: PTFE, HV/UHV: fiberglass |
Coarse Positioning Mode | |
travel range (step mode) | 10 ° |
maximum drive velocity @ 300 K | 3 °/s |
input voltage range | 0 - 45 V |
Fine Positioning Mode | |
fine angular positioning range @ 300 K | 1.1 m° |
fine positioning resolution | µ° |
input DC voltage range | 0 - 45 V |
Position Encoder | |
readout mechanism | optoelectronic sensor |
encoded travel range | entire travel |
sensor resolution | 1 µ° |
sensor power (when measuring) | 300 mW |
repeatability | 50 µ° (bidirectional) |
Load (@ ambient conditions) | |
maximum load | 1 kg |
maximum dynamic torque around axis | 43.5 Ncm |
General Specifications | |
environment | /RT |
operating temperature range | 17 - 25°C |
operating humidity range | 40 - 60% |