ANSx150 LT x Scanner aus Titan Nanopositioners for Low Temperature
Icon Load

100 g

Icon Room Temperature

1 .. 373 K

Icon Pressure

ambient

Icon Magnetic Fields

0 - 35 T

ANSx150/LT x-Scanner made from Titanium

large range open loop x-scanner

This piezo driven scanner is optimized for high precision scanning in one direction over a large scan range, even at cryogenic temperatures.
The positioner is designed for temperatures between 1K..373 K.
Titanium is the material of choice for our cryogenic nanopositioners, as it is light, nonmagnetic, and has a coefficient of expansion matching the piezo and other elements of the drive for reliable low temperature performance.
Cryogenic Tablet

Specifications

For a detailed definition of used terms and descriptions please visit our Nanopositioners Glossary

Size and Dimensions
footprint; height24 mm x 24 mm; 9 mm
maximum installation space24 mm x 24 mm; 9 mm
weight18 g
Materials
positioner bodytitanium
actuatorPZT ceramics
connecting wiresinsulated twisted pair, copper
Fine Positioning Mode
fine linear positioning range @ 300 K80 µm
fine linear positioning range @ 4 K125 µm
fine positioning resolutionsub-nm
typical actuator capacitance fine @ 300 K11.1 µF
typical actuator capacitance fine @ 4 K2.6 µF
input DC voltage range @ 300 K0 - 60 V
input DC voltage range @ 4 K+/- 120V
*scan range depends on environment
Load (@ ambient conditions)
maximum load100 g
General Specifications
environment/LT