ANGp101 RES LT UHV   Goniometer zur Phi Verkippung Nanopositioners for Low Temperature
Icon Load

100 g

Icon Travel Range Gonio

5.4 °

Icon Room Temperature

1 .. 423 K

Icon Pressure

5E-11 mbar

Icon Magnetic Fields

0 - 35 T

Icon Resolution

0.1 m°

ANGp101/RES/LT/UHV - goniometer for "phi" tilting

goniometer for Φ-positioning

The ANGp101 provides high resolution Φ - positioning for 2" diameter setups. Combined with the ANGt101, the ANGp101 forms the lower positioner of a powerful Euler Goniometer with a common center of rotation. The small footprint and titanium body make it particularly suited for compact setups and magnetically sensitive applications.
The positioner is designed for temperatures between 1K..423 K and pressures down to 5 E-11 mbar (/LT/UHV), all positioners can be baked out up to 150°C .
Titanium is the material of choice for our cryogenic nanopositioners, as it is light, nonmagnetic, and has a coefficient of expansion matching the piezo and other elements of the drive for reliable low temperature performance.
Cryogenic Tablet

Specifications

For a detailed definition of used terms and descriptions please visit our Nanopositioners Glossary

Size and Dimensions
footprint; height24 mm x 24 mm; 11 mm
distance center of rotation to top51 mm
weight18 g
height11 mm
Materials
positioner bodytitanium
actuatorPZT ceramics
connecting wiresinsulated twisted pair, copper
Coarse Positioning Mode
travel range (step mode)5.4 °
maximum drive velocity @ 300 K~ 1 °/s
input voltage range0 - 60 V
Fine Positioning Mode
fine positioning resolutionno fine positioning capability
Position Encoder
readout mechanismresistive sensor
encoded travel rangefull travel
sensor resolution0.1 m°
sensor power (when measuring)0.01 - 1 mW
repeatability2 m°
Load (@ ambient conditions)
maximum load100 g
maximum dynamic force along the axis2 N
General Specifications
environment/LT/UHV
show more icon Download PDF