IDSH M12 F40 LT UHV RAD IDSH

IDSH/M12/F40/LT/UHV/RAD

Ø 14 mm, length 19.5 mm

The focusing sensor head M12/F40 is used for standard applications, e.g. for measuring vibrations or material deformations. Because of the larger lense's diameter, it enables a quite high angular tolerance of ±0.35°. The housing is designed for high stability and the sensor head is capable of measuring on various applications.
The sensor head is designed for temperatures between mK..423 K, pressures down to 5 E-11 mbar and radiation harsh environments. The sensor kit package includes two glass fibers. A LT/UHV/RAD compatible glass fiber (length: 2m) connecting the sensor head to an optical feedthrough, and an ambient glass fiber (length: 5 m) connecting the feedthrough to the interferometer controller.
Cryogenic Tablet

Specifications

Sensor Head
optics typefocusing
dimensionsØ 14 mm, length 19.5 mm
weight [g]5.1
mountingmetric M12 x 0.5
focal length [mm]40
typical beam diameter at focal point [µm]~ 45
connectorFC/PC
working range (target: glass) [mm]38.5 .. 41.5
angular tolerance (target: glass) [°]± 0.35
working range (target: mirror) [mm]46 .. 50
angular tolerance (target: mirror) [°]± 0.35
fiber connectionscrewed
Environment
pressure1E-10 mbar .. 5 bar
temperature0 .. 150°C
Main Materials
housingTitanium Grade 2 (3.7035)
lensglass (with AR C coating)
adhesiveNASA outgassing compliant adhesive
Scope of Delivery
fiber IDS-feedthroughIDSMF/APC/RT (1009981)
fiber feedthrough-sensor headIDSMF/UHV metal tubing (1016354)
number of axes1
mirrorØ 9 mm, Al coated (1010846)
show more icon Download PDF